WDXRF on Your Wavelength Episode #7 Recap
Nov 11, 2026
Recap provided by Bridget Marriott
How fine should samples be ground? Are matrix-matched standards always necessary for fused bead calibrations? Can XRF determine what lies beneath gold plating? And what should laboratories consider when investing in a new instrument?
These are just some of the topics explored during episode 7 of WDXRF on Your Wavelength, on November 6, 2026, which features insights from Bob Burton, XRF Sales Manager at Rigaku. The discussion covered sample preparation, calibration strategies, instrument selection, and analytical performance, while also addressing audience questions on LOI, gold plating analysis, and calibration best practices.
Whether you're new to WDXRF or have years of experience using the technique, the questions and answers below provide practical guidance that can help improve analytical results and support more informed decision-making. If you missed the live event, you can watch the full recording here.
Episode Recap
-
One of the biggest challenges is overcoming the perception that XRF is still a highly complex technique that requires a dedicated expert to operate. Decades ago, XRF systems were complicated and often required a specialist whose primary responsibility was running the instrument.
Today, modern WDXRF systems are much more user-friendly, with advanced software and built-in intelligence that guide users through the process. The instruments are designed to work for the user, not the other way around. The biggest hurdle is helping users become comfortable with the technology and trust that modern systems can handle much of the complexity behind the scenes.
-
Regardless of the application, sample preparation and calibration consistently create the most challenges.
Sample preparation directly affects data quality. If samples are not prepared properly, analytical results will suffer. For powdered samples, grinding to approximately 50 microns is often recommended. A useful way to think about this particle size is that properly ground powders typically have a smooth consistency similar to talc or cornstarch rather than a gritty texture.
Calibration is equally important because WDXRF accuracy depends on the quality of the standards used to build the calibration. The instrument itself delivers excellent precision and stability, but reliable standards and sound calibration practices are what determine analytical accuracy.
-
These terms are often used interchangeably, but they describe different aspects of analytical performance.
Accuracy refers to how close a measurement is to the true or accepted value and depends heavily on calibration standards.
Precision describes how consistently the instrument can generate similar results under the same conditions.
Repeatability measures how closely results agree when the same sample is analyzed multiple times without changing anything.
Reproducibility evaluates the variation observed when multiple samples are prepared and analyzed, potentially by different operators. This makes reproducibility a measure of the entire process, including sample preparation, operator technique, and instrument performance.
WDXRF instruments are inherently very precise and repeatable. Accuracy, however, is primarily driven by the quality of the calibration and standards.
-
XRF is used in a surprisingly wide range of industries, often in ways that many people would never expect. One memorable example is the use of XRF to test edible underwear for heavy metal contamination.
While unusual, it highlights one of XRF's greatest strengths: versatility. From mining and cement production to food products and consumer goods, XRF can be adapted to analyze an enormous variety of materials. The key challenge is often determining the best way to prepare and present the sample for analysis. Once that is solved, the range of potential applications is remarkably broad.
-
When available, matrix-matched standards generally provide the best calibration performance because they most closely resemble the samples being analyzed.
However, fused bead preparation significantly reduces matrix effects by dissolving samples into a consistent glass matrix, making calibration more straightforward than many other XRF methods. Because of this, both matrix-matched standards and calibrations developed from high-purity materials can be effective approaches.
Success ultimately depends on factors such as proper sample preparation, homogeneity, accurate calculations, and verification of the calibration's performance. Selecting the appropriate calibration materials is important, but building a reliable calibration requires attention to the entire analytical process.
-
Gold presents a unique challenge because it absorbs X-rays extremely efficiently.
Once a gold layer becomes sufficiently thick, it can prevent X-rays from reaching and returning from the material underneath. In those situations, XRF can confirm the composition of the gold surface but may not be able to determine what lies beneath the plating without resorting to destructive testing methods.
This limitation is one reason authenticating gold objects remains a challenging application for non-destructive analysis.
-
A question that is often overlooked is: "What will I want this instrument to do five or ten years from now?"
Many purchasing decisions focus on solving today's analytical challenges, but laboratory needs rarely stay the same. New products, changing regulations, evolving customer requirements, and future applications can all create demands that were not part of the original purchasing decision.
Considering long-term needs during instrument selection can help ensure the system remains valuable for years to come and may prevent costly upgrades or limitations down the road.
-
Support should be one of the most important considerations.
Analytical performance is important, but any instrument is only valuable when it is operating and producing results. Access to knowledgeable technical support, application assistance, local service resources, and replacement parts can have a major impact on instrument uptime and overall productivity.
For many laboratories, the quality of support can be just as important as the specifications listed in a brochure.
-
Glass bead fusion is commonly used to minimize matrix effects and improve analytical consistency across different sample types.
During the fusion process, the sample is dissolved in a borate flux, creating a uniform glass bead. Because the resulting matrix is much more consistent from sample to sample, calibrations are often easier to develop and can be applied across a wider range of materials.
This approach is particularly effective for major and minor elemental analysis. However, because fusion dilutes the sample, laboratories often use pressed pellets instead when trace-element sensitivity is the primary concern. Choosing between fused beads and pressed pellets depends on the analytical objectives and the concentrations being measured.
-
Loss on ignition (LOI) can have a significant impact on analytical results and should be considered during calibration development, particularly for fused bead applications.
Moisture and volatile components contribute to sample mass but not to the elemental concentrations being measured. For this reason, materials should be properly dried when necessary, and consistent preparation procedures should be followed throughout calibration development and routine analysis.
More broadly, controlling sources of variability such as moisture content, grinding time, binder ratios, and sample weights is essential for achieving reliable and repeatable results.
Even small inconsistencies in sample preparation can introduce error, making standardized procedures an important part of achieving high-quality analytical results.
Join us next time!
WDXRF on Your Wavelength will be back every month with more answers, tips, and insights from our team. If you use WDXRF or are curious about it, we hope you’ll join us live next time!
The live event information is posted on our LinkedIn event page.
Do you have questions in the meantime? Drop your question in the comment section of the most recent episode. Our team will either respond directly or address your question live during the next episode.