NEX QC—CCA Wood Treatment
Scope
The measurement of chromated copper arsenate (CCA) treated wood and wood treatment solutions is demonstrated.
Background
Wood treatments are used to protect lumber from fungi, insects, UV damage and general wear. In recent years, the US EPA has asked for voluntary compliance to restrict the use of CCA due to concerns about chromium and arsenic leaching into soils. Lumber treated with CCA is used in a limited number of situations, such as in roofing shingles, permanent building foundations, and certain commercial applications. When treating wood, the proper balance of treatment solution must be monitored to ensure the highest quality while minimizing waste and excess cost due to treatment usage or product rejection. Chromium, copper and arsenic levels are monitored in solution prior to treatment, and then in the wood to ensure proper retention. A quick, simple, reliable means of analysis is required throughout the quality control process. XRF is an ideal tool for such analysis.
Model: NEX QC
Calibration – CCA in wood
An empirical calibration was built using a set of assayed wood standards.
Element: CrO₃ Units: % |
||
Sample I.D. | Standard value | Calculated value |
W-C | 0.46 | 0.464 |
W-D | 0.62 | 0.612 |
W-E | 0.86 | 0.844 |
W-G | 1.52 | 1.560 |
W-H | 2.05 | 2.058 |
W-I | 3.00 | 2.951 |
W-J | 4.15 | 4.164 |
Calibration plot CrO₃ in wood
Element: CuO Units: % |
||
Sample I.D. | Standard value | Calculated value |
W-C | 0.18 | 0.185 |
W-D | 0.24 | 0.237 |
W-E | 0.32 | 0.309 |
W-G | 0.56 | 0.591 |
W-H | 0.74 | 0.726 |
W-I | 1.10 | 1.083 |
W-J | 1.59 | 1.599 |
Calibration plot CuO in wood
Element: As₂O₅ Units: % |
||
Sample I.D. | Standard Value | Calculated Value |
W-C | 0.33 | 0.328 |
W-D | 0.43 | 0.427 |
W-E | 0.58 | 0.586 |
W-G | 1.04 | 1.052 |
W-H | 1.42 | 1.420 |
W-I | 2.13 | 2.045 |
W-J | 3.02 | 3.024 |
Calibration plot As₂O₅ in wood
Repeatability – CCA in wood
To demonstrate repeatability (precision), the low and high samples were chosen from the set of calibration standards. Each sample was measured in static position.
Element: Cr₂O₃ Units: % |
||||
Sample I.D. | Standard Value | Average Value | Std. Dev | % Relative |
W-C | 0.46 | 0.463 | 0.004 | 0.9 |
W-J | 4.15 | 4.155 | 0.011 | 0.6 |
Element: CuO Units: % |
||||
Sample I.D. | Standard Value | Average Value | Std. Dev | % Relative |
W-C | 0.18 | 0.187 | 0.001 | 0.5 |
W-J | 1.59 | 1.653 | 0.012 | 0.8 |
Element: As₂O₅ Units: % |
||||
Sample I.D. | Standard Value | Average Value | Std. Dev | % Relative |
W-C | 0.33 | 0.331 | 0.001 | 0.3 |
W-J | 3.02 | 3.044 | 0.017 | 0.6 |
Calibration – CCA in solution
An empirical calibration was built using a set of assayed solution standards.
Element: CrO₃ Units: % |
||
Sample I.D. | Assay value | Calculated value |
S-A | 2.859 | 2.862 |
S-C | 3.099 | 3.118 |
S-D | 3.215 | 3.197 |
S-E | 3.336 | 3.355 |
S-G | 3.573 | 3.524 |
S-H | 3.811 | 3.835 |
Calibration plot CrO₃ in solution
Element: CuO Units: % |
||
Sample I.D. | Assay value | Calculated value |
S-A | 1.050 | 1.056 |
S-C | 1.139 | 1.145 |
S-D | 1.182 | 1.182 |
S-E | 1.266 | 1.246 |
S-G | 1.313 | 1.310 |
S-H | 1.400 | 1.411 |
Calibration plot CuO in solution
Element: As₂O₅ Units: % |
||
Sample I.D. | Assay value | Calculated value |
S-A | 2.082 | 2.0770 |
S-C | 2.257 | 2.2640 |
S-D | 2.342 | 2.3320 |
S-E | 2.430 | 2.4560 |
S-G | 2.603 | 2.6030 |
S-H | 2.755 | 2.7570 |
Calibration plot As₂O₅ in solution
Repeatability – CCA in solution
To demonstrate repeatability (precision), the low and high samples were chosen from the set of calibration standards. Each sample was measured in static position.
Element: Cr₂O₃ Units: % |
||||
Sample I.D. | Standard value | Average value | Std. dev | % Relative |
S-A | 2.859 | 2.855 | 0.0173 | 0.6 |
S-H | 3.811 | 3.818 | 0.0229 | 0.6 |
Element: CuO Units: % |
||||
Sample I.D. | Standard value | Average value | Std. dev | % Relative |
S-A | 1.050 | 1.057 | 0.0033 | 0.3 |
S-H | 1.400 | 1.409 | 0.0068 | 0.5 |
Element: As₂O₅ Units: % |
||||
Sample I.D. | Standard value | Average value | Std. dev | % Relative |
S-A | 2.082 | 2.078 | 0.0174 | 0.8 |
S-H | 2.755 | 2.751 | 0.0161 | 0.6 |
Retention report
To measure a wood sample, enter the density of the wood. The measurement calculates concentrations of Cr₂O₃, CuO and As₂O₅ and also outputs balance and retention values.
Conclusion
The typical results detailed here show exceptional performance for the fast and simple measurement of CCA in wood and solution. The Rigaku NEX QC is an excellent tool along the QC process in producing treated lumber, giving the production process an affordable means of optimizing quality while minimizing costs and helping to minimize product rejection and waste.