Application Note BATT1011
Introduction
In XRD using Cu radiation sources, peak profile backgrounds for cathode materials are generally high due to the effects of the transition metal elements they contain. This makes the detection of trace crystal phase peaks difficult. The XSPA-400 ER uses high energy resolution to decrease X-ray fluorescence originating from samples and reduce background components to achieve higher-sensitivity measurements compared to traditional detectors.
Crystal phase analysis
- Analysis: Cathode material
- Analysis method: Qualitative analysis
- Use: Optimizing electrochemical performance
- Analyzed materials: Li(NixCoyMnz)O₂, NCM
Figure 1: Profile measured using traditional detector and XSPA-400 ER
Figure 2: Magnified XRD profile measured using traditional detector and XSPA-400 ER
Conclusion
The XSPA-400 ER successfully lowers background compared to traditional detectors. As such, it makes it easier to observe minute peaks, which means trace crystal phases can be observed.