Application Note EDXRF3289

Scope
The analysis of phosphorus (P) and sulfur (S) in coal is demonstrated using the Rigaku NEX QC II+ energy dispersive X-ray fluorescence (EDXRF) analyzer.
Background
Coke used in the steelmaking process is required to have a low phosphorus (P) content. P is a critical impurity that degrades quality and therefore is considered a penalty element and must be below the maximum limit of 0.05%. Similarly, S is a penalty element for emissions and environmental reasons. The P and S content in coke is directly related to the P and S content of the parent coal, and so P and S concentrations are used in part to grade coking coal. Therefore, a fast and reliable means of measuring and monitoring P and total S content is vital for the grading and processing of coking coal. To meet industry needs, Rigaku offers the NEX QC II+, a simple and versatile benchtop EDXRF analyzer for the analysis of phosphorus, sulfur, and other elements in coal.
Calibration
An empirical calibration was built for P and total S using a set of 15 assayed coal powder samples provided by a coal processing facility. Aluminum (Al), silicon (Si), calcium (Ca), titanium (Ti), iron (Fe), and strontium (Sr) were also measured to enable absorption/enhancement corrections (alpha corrections) that compensate for variations in matrix X-ray absorption/enhancement effects.
| Element: P Units: mass % |
||
| Sample I.D. | Standard value | Calculated value |
| 1 | 0.028 | 0.0289 |
| 2 | 0.035 | 0.0346 |
| 3 | 0.024 | 0.0222 |
| 4 | 0.017 | 0.0192 |
| 5 | 0.023 | 0.0220 |
| 6 | 0.025 | 0.0255 |
| 7 | 0.027 | 0.0289 |
| 8 | 0.030 | 0.0276 |
| 9 | 0.044 | 0.0449 |
| 10 | 0.049 | 0.0481 |
| 11 | 0.061 | 0.0610 |
| 12 | 0.061 | 0.0607 |
| 13 | 0.068 | 0.0654 |
| 14 | 0.064 | 0.0644 |
| 15 | 0.068 | 0.0680 |

Correlation plot P
| Element: S Units: mass % |
||
| Sample I.D. | Standard value | Calculated value |
| 4 | 0.51 | 0.510 |
| 5 | 0.55 | 0.549 |
| 6 | 0.61 | 0.611 |
| 7 | 0.60 | 0.603 |
| 8 | 0.61 | 0.607 |
| 9 | 0.39 | 0.390 |
| 10 | 0.50 | 0.500 |
| 11 | 0.28 | 0.281 |
| 12 | 0.41 | 0.410 |
| 13 | 0.31 | 0.309 |
| 14 | 0.27 | 0.269 |
| 15 | 0.35 | 0.350 |
Correlation plot S
Repeatability
Representative samples from the calibration suite were selected to demonstrate typical instrument repeatability (precision). Ten repeat analyses of each sample were performed with the sample in a static position.
| Element: Std 5 Units: mass % |
||||
| Element | Standard value | Average value | Std. Dev | % Relative Dev |
| P | 0.023 | 0.0222 | 0.0002 | 0.9 |
| S | 0.55 | 0.5533 | 0.0027 | 0.5 |
| Element: Std 14 Units: mass % |
||||
| Element | Standard value | Average value | Std. Dev | % Relative Dev |
| P | 0.064 | 0.0645 | 0.0002 | 0.3 |
| S | 0.27 | 0.2659 | 0.0018 | 0.7 |
Conclusion
The results presented here demonstrate the ability of the NEX QC II+ to reliably measure phosphorus and sulfur in coal, supporting the quality control requirements for low contaminant levels in steelmaking. With its intuitive, icon-driven touchscreen interface and embedded computer, the NEX QC II+ simplifies routine analysis and operator training without requiring an external PC. Its compact design and minimal maintenance requirements make it a practical solution for quality control applications throughout the coal and steelmaking industries.