Pb in Gasoline by ASTM D5059 Part C and A
Scope
Cartesian Geometry EDXRF performance is demonstrated for the measurement of lead (Pb) in gasoline as per ASTM D5059 using the Bi internal standard method Part C (ultra-low Pb) and Part A (high Pb).
Background
Tetraethyl lead (TEL) is an anti-knock agent added to gasoline, or petrol. Lead is a toxic metal that interferes with anti-pollution devices and contributes to lead poisoning. In the 1970s, regions around the world began phasing out lead in MoGas (motor gasoline), beginning in the United States and Europe. Lead is still used in AvGas (aviation gasoline) for some piston engine propeller aircraft; however, efforts are being made to lower and remove lead in this fuel as well. One of the main international standard test methods for the measurement of lead in gasoline is ASTM D5059 using X-ray spectroscopy, XRF (X-ray fluorescence). As efforts continue to reduce and remove lead from gasoline fuels worldwide, Applied Rigaku Technologies offers EDXRF analyzer NEX CG II using indirect excitation Cartesian Geometry for the measurement of lead by ASTM D5059 as well as many other applications in the petroleum and fuel industries.
Model: NEX CG II
Calibration
Empirical calibration was made using commercially available standards. The user spikes each standard and the blank with the Bi IS as described for Part C or A. Calibration is then g/US gal Pb vs. I, where I is the X-ray intensity expressed as (Pb net cps / Bi gross cps).
Part C
Element: Pb Units: g/US gal |
||
Sample I.D. | Standard value | Calculated value |
1 | 0.001 | 0.0007 |
2 | 0.005 | 0.0051 |
3 | 0.010 | 0.0103 |
4 | 0.050 | 0.0505 |
5 | 0.100 | 0.9960 |
6 | 0.300 | 0.3000 |
Correlation plot Pb Part C
Part A
Element: Pb Units: g/US gal |
||
Sample I.D. | Standard value | Calculated value |
1 | 0.1 | 0.108 |
2 | 1.0 | 0.990 |
3 | 2.0 | 1.986 |
4 | 3.0 | 3.020 |
5 | 4.0 | 4.005 |
6 | 5.0 | 4.991 |
Correlation plot Pb Part A
Precision
Instrument repeatability (precision) is determined by five repeat analyses of each sample in static position. To minimize evaporation error, five repeat measurements of each sample were taken for demonstration.
Element: Pb Units: g/US gal |
||||
Sample | Assay value | Average measured value | Std. dev | % Relative deviation |
Part C 2 | 0.0050 | 0.0054 | 0.0006 | 12% |
Part C 6 | 0.3000 | 0.3008 | 0.0015 | 0.5% |
Part A 1 | 0.1000 | 0.1010 | 0.0015 | 1.5% |
Part A 2 | 1.000 | 0.9921 | 0.0013 | 0.1% |
Part A 6 | 5.000 | 4.932 | 0.0067 | 0.1% |
ASTM D5059
Standard Test Methods for Lead in Gasoline by X-Ray Spectroscopy
The test methods cover the determination of the total lead content of gasoline within the following concentration ranges:
Part C | Part A |
0.010 to 0.5 g Pb/US gal | 0.10 to 5.0 g Pb/US gal |
0.012 to 0.6 g Pb/UK gal | 0.12 to 6.0 g Pb/UK gal |
0.0026 to 0.132 g Pb/L | 0.026 to 1.32 g Pb/L |
Conversion factors: 3.7854 g/US gal = 4.5461 g/UK gal = 1.0000 g/L
Conclusion
The results shown here indicate NEX CG II easily meets ASTM D5059 Parts A and C using the Bi internal standard method for the measurement of lead in gasoline.