NEX QC—Plating Baths
Scope
The analysis of various plating baths is demonstrated for electroless and electroplating applications.
Background
Metal parts are often plated to create a decorative surface finish or to ensure desired physical properties, such as corrosion and salt resistance, lubricity, protection from wear and abrasion, and to improve the hardness of the plated metal. Electro-plating and electroless plating are commonly used in almost every industry, from aerospace and electronics to the gas and oil industry and general construction. Monitoring the bath composition is vital in quality control to ensure finished parts are plated to the correct thickness to achieve desired properties without over-coating and wasting chemical materials. Applied Rigaku Technologies offers NEX QC benchtop EDXRF analyzer for the measurement of plating bath composition in a small, easy to use tool, ideal for at-line quality checks or quality confirmation in the lab.
Model: NEX QC
Units
1 oz/gal = 7.489 g/L
1 g/L = 0.134 oz/gal
Ni Bath
Calibration
Element: Ni Units: oz/gal |
||
Sample I.D. | Assay value | Calculated value |
1 | 7.773 | 7.788 |
2 | 10.498 | 10.373 |
3 | 12.250 | 12.213 |
4 | 13.719 | 13.592 |
5 | 15.395 | 15.236 |
Correlation plot Ni bath
Recovery and repeatability
To demonstrate recovery and repeatability (precision), low and high calibration standards were analyzed in a static position for 10 consecutive measurements.
Sample ID | Units | Assay value | NEX QC avg. value | Standard deviation | RSD % |
Std 1 | oz/gal | 7.773 | 7.742 | 0.037 | 0.5 |
Std 5 | oz/gal | 15.395 | 15.425 | 0.043 | 0.3 |
CuCN Bath
Calibration
Element: CuCN Units: oz/gal |
||
Sample I.D. | Assay value | Calculated value |
1 | 2.500 | 2.518 |
2 | 3.457 | 3.464 |
3 | 4.800 | 4.710 |
4 | 5.545 | 5.627 |
5 | 6.500 | 6.483 |
Correlation plot CuCn bath
Recovery and repeatability
To demonstrate recovery and repeatability (precision), low and high calibration standards were analyzed in a static position for 10 consecutive measurements.
Sample ID | Units | Assay value | NEX QC avg. value | Standard deviation | RSD % |
Std 1 | oz/gal | 2.500 | 2.530 | 0.010 | 0.4 |
Std 5 | oz/gal | 6.500 | 6.531 | 0.015 | 0.2 |
AgCN Bath
Calibration
Element: AgCN Units: oz/gal |
||
Sample I.D. | Assay value | Calculated value |
1 | 0.100 | 0.099 |
2 | 1.000 | 1.000 |
3 | 2.000 | 2.011 |
4 | 3.200 | 3.179 |
5 | 4.000 | 4.011 |
Correlation plot AgCN bath
Recovery and repeatability
To demonstrate recovery and repeatability (precision), low and high calibration standards were analyzed in a static position for 10 consecutive measurements.
Sample ID | Units | Assay value | NEX QC avg. value | Standard deviation | RSD % |
Std 1 | oz/gal | 0.100 | 0.100 | 0.0004 | 0.4 |
Std 5 | oz/gal | 4.000 | 4.036 | 0.008 | 0.2 |
CdCN Bath
Calibration
Element: CdCN Units: oz/gal |
||
Sample I.D. | Assay value | Calculated value |
1 | 1.500 | 1.500 |
2 | 2.905 | 2.911 |
3 | 4.300 | 4.469 |
4 | 5.597 | 5.636 |
5 | 7.000 | 6.986 |
Calibration plot CdCN bath
Recovery and repeatability
To demonstrate recovery and repeatability (precision), low and high calibration standards were analyzed in a static position for 10 consecutive measurements.
Sample ID | Units | Assay value | NEX QC avg. value | Standard deviation | RSD % |
Std 1 | oz/gal | 1.500 | 1.500 | 0.005 | 0.3 |
Std 5 | oz/gal | 7.000 | 6.997 | 0.033 | 0.5 |
Conclusion
The results indicate NEX QC system yields exceptional performance for the analysis of plating baths. The rugged self-contained system with simple, intuitive software makes the NEX QC analyzer an ideal tool for at-line quality control or in the QC lab itself. This methodology can also be adapted for in-line process control in real time with the Rigaku NEX OL process analyzer.