NEX QC—Penta Wood Treatment
Scope
The measurement of pentachlorophenol (Penta) in treated wood and wood treatment solutions is demonstrated.
Background
Wood treatments are used to protect lumber from fungi, insects, UV damage, water damage, and general wear. Pentachlorophenol (called Penta or PCP) is typically not used to treat wood used for residential application. It is used for heavier more industrial applications like treating telephone poles or materials for bridges. When treating wood, the proper balance of treatment solution must be monitored to ensure the highest quality while minimizing waste and excess cost due to treatment usage or product rejection. Penta levels are monitored in solution prior to treatment, and then in the wood to ensure proper retention. A quick, simple, reliable means of analysis is required throughout the quality control process. XRF is an ideal tool for such analysis.
Model: NEX QC
Calibration – Penta in wood
An empirical calibration was built using a set of assayed wood standards.
Penta Units: % |
||
Sample I.D. | Standard value | Calculated value |
W-A | 0.998 | 0.991 |
W-B | 1.350 | 1.369 |
W-C | 1.470 | 1.457 |
W-D | 2.090 | 2.090 |
Correlation plot Penta in wood
Repeatability – Penta in wood
To demonstrate repeatability (precision), the low and high samples were chosen from the set of calibration standards. Each sample was measured in static position.
Penta Units: % |
||||
Sample I.D. | Standard value | Average value | Std. dev | % Relative |
W-A | 0.998 | 0.994 | 0.004 | 0.4 |
W-D | 2.090 | 2.094 | 0.006 | 0.3 |
Calibration – Penta in solution
An empirical calibration was built using a set of assayed solution standards.
Penta Units: % |
||
Sample I.D. | Standard value | Calculated value |
S-A | 0.505 | 0.509 |
S-B | 1.019 | 0.962 |
S-C | 1.510 | 1.503 |
S-D | 1.998 | 2.008 |
S-E | 2.504 | 2.510 |
S-F | 3.009 | 3.038 |
S-G | 5.007 | 4.996 |
S-H | 7.506 | 7.505 |
Correlation plot Penta in solution
Repeatability – Penta in solution
To demonstrate repeatability (precision), the low and high samples were chosen from the set of calibration standards. Each sample was measured in static position.
Penta Units: % |
||||
Sample I.D. | Standard value | Average value | Std. dev | % Relative |
S-A | 0.505 | 0.509 | 0.003 | 0.6 |
S-H | 7.506 | 7.500 | 0.020 | 0.3 |
Conclusion
The typical results detailed here show exceptional performance for the fast and simple measurement of Penta in wood and solution. The Rigaku NEX QC is an excellent tool along the QC process in producing treated lumber, giving the production process an affordable means of optimizing quality while minimizing costs and helping to minimize product rejection and waste.