Master Thermal Analysis in Just 30 Minutes!
2. Feeling the Heat: What Can You Do with DSC

This is a written summary of a live webinar presented on February 26, 2026. The recording and resources are available on the recording page.
Presented by:
Jing Sun
Thermal Analysis Group
Rigaku
Webinar summary
Differential scanning calorimetry, or DSC, is a thermal analysis technique where the difference between the heat-flow rates into a sample and a reference material is measured during controlled heating or cooling. It can detect and quantify thermal events such as melting, crystallization, glass transition, decomposition, oxidation, and other reactions. A stable region with no heat-flow rate difference is observed between sample and reference forms the baseline. In the described convention, downward peaks represent endothermic events, while upward peaks represent exothermic events. Unlike DTA, DSC can measure reaction enthalpy, making it useful for quantitative analysis of heat flow, enthalpy change, heat capacity, and reaction temperatures.
DSC instruments are generally divided into power-compensation and heat flux types. Heat-flux DSC places the sample and reference material within symmetrically an electric furnace, with thermocouples beneath the sensor plate detecting the temperature difference between them. During typical heating, both the sample and reference material follow the furnace temperature with a slight delay. When the sample undergoes an endothermic process, the sample temperature absorbs heat and temporarily lags behind the reference temperature, producing an endothermic peak. During an exothermic process, heat from the reaction increases the sample temperature relative to the reference, producing an exothermic peak. Heat flux DSC is described as offering high sensitivity and a stable baseline, while power-compensation DSC is described as more susceptible to baseline drift and noise.
Obtaining reliable DSC data requires proper calibration, careful sample preparation, and appropriate measurement conditions. Instrument response characteristics may change over operational time because of transducer aging, oxide accumulation, electronic drift, or contamination from sample spills. Regular baseline, temperature, and energy checks are therefore indispensable. Calibration using multiple temperature or energy standards is recommended, and indium is used as an example to verify temperature accuracy and enthalpy reproducibility.
Sample pan or crucible selection has a major effect on data quality. The sample pan must be compatible with the sample and the temperature range. Aluminum pans are suitable up to about 500°C, while alumina pans can be used up to about 1500°C. Quartz pans are useful for measuring liquid in open pans up to about 1000°C, stainless-steel pans can be used for high-pressure, sealed measurements up to 5 MPa, and platinum pans offer excellent thermal conductivity up to about 1500°C but may cause catalytic reactions with some samples. For example, indium measured in an aluminum pan showed higher intensity and sensitivity than measurements performed in an alumina pan due to aluminum’s higher thermal conductivity.
Sample preparation should maximize thermal contact with the sample pan and avoid contamination. Powders should be spread thinly and uniformly and gently pressed against the bottom of the pan. Blocks or pellets should be cut to expose a large flat, smooth surface as possible. Liquids should be dispensed into the bottom center of the pan without spilling into the sealing area. Sheet or film samples should be cut to fit the inner diameter of the pan, and several layers may be stacked if more mass is needed. The outside surface of the pan should be checked for contamination before measurement because any residue could contaminate the measuring instrument and cause measurement errors.
Sample mass affects both data sensitivity and peak resolution. Larger samples increase sensitivity and produce intense signals, but the temperature distribution within the sample becomes more pronounced, which can lead to broader peak widths. Additionally, because the reaction process takes longer, the apparent reaction temperature may shift toward higher temperatures. Smaller samples improve peak resolution by narrowing transition ranges, reducing the overlap of nearby thermal events. The recommended approach is to use the smallest sample mass while still achieving sufficient sensitivity. Although the reference mass may shift the baseline due to heat-capacity differences compared to the sample, but it does not change the sample’s actual thermal behavior as long as the reference remains thermally inert across the full temperature range.
Measurement conditions also significantly influence DSC results. The temperature range should begin at least 20–30°C below the expected transition temperatures to establish a stable baseline, especially for relatively weak signals such as glass transitions. The upper temperature limit should be set high enough to capture the thermal event of interest but low enough to avoid unwanted decomposition. For unknown materials, performing preliminary STA measurement before selecting a DSC measurement temperature range can help identify the temperature at which mass loss begins. Heating rate involves a trade-off: higher heating increases sensitivity and shortens the measurement time, but shifts the onset, peak, and offset temperatures higher value; lower heating improves peak resolution and makes it easier separate overlapping thermal phenomena. Measurement atmosphere selection is equally important. Air atmosphere promotes oxidation and combustion, while using inert gas such as nitrogen allows for the observation of thermal behavior without oxidation. Dedicated DSC models cannot be used at temperatures where the sample decomposes because the sensors are easily contaminated and damaged. For example, with Polyethylene, , STA reveals that while exothermic combustion resulting in mass loss occurs at approximately 200°C in air, endothermic decomposition resulting in mass loss occurs at temperatures above 400°C in nitrogen.
Several application examples are presented to illustrate what DSC can reveal. Anhydrous caffeine showed a phase transition near 160°C and a sharp melting peak around 237°C. Grinding caused the transition temperature to decrease from Form II to Form I; since the transition temperature dropped to approximately 150°C after just 3 minutes of grinding, it was demonstrated that mechanical processing can alter the thermal properties of solids. This is known as the mechanochemical effect. Amorphous indomethacin showed changes during storage, including shifts in glass transition temperature and the appearance of crystallization peaks. These results show how DSC can monitor issues related to physical stability, recrystallization, and potential shelf-life in pharmaceutical materials. Carbamazepine displayed melting, recrystallization, and subsequent melting of a more stable phase or another polymorph, demonstrating the value of combining DSC data with visual sample observation capabilities. A lithium-ion battery separator showed two melting peaks, at 132°C and 166°C, corresponding to different polymer layers; the lower-melting PE layer supports thermal shutdown behavior, an important battery safety function.
The latest DSC systems may feature high sensitivity, wide temperature range, self-diagnostic capabilities, sample observation cameras, cooling units, and automatic sample changers. Sample observation capabilities allow for real-time visual monitoring of shape, color, and physical state, enabling the interpretation of both visible sample changes and thermal data. Automated sample handling supports high-throughput workflows involving numerous samples and reference materials.
Frequently asked questions
-
DSC measures the difference in heat flow or temperature behavior between a sample and a reference while both are heated or cooled under controlled conditions. It is used to identify and quantify thermal events such as melting, crystallization, glass transitions, decomposition, oxidation, and other reactions. It can also provide quantitative information such as reaction enthalpy, heat capacity, and reaction temperature.
-
An endothermic event occurs when the sample absorbs heat, such as during melting. In the described DSC curve convention, this appears as a downward peak. An exothermic event occurs when the sample releases heat, such as during crystallization, oxidation, or combustion. In the described convention, this appears as an upward peak.
-
Calibration is needed because an instrument’s response can change over time because of sensor aging, oxide buildup, electronic drift, sample spills, or cleaning. These changes can affect the baseline, temperature accuracy, and measured energy values. Regular checks of baseline drift, temperature, and energy help ensure reliable results. If measurements fall outside laboratory tolerances, calibration should be performed.
-
The pan should be selected based on the sample chemistry, measurement temperature, atmosphere, and pressure requirements. Aluminum pans are useful for many lower-temperature measurements, alumina and platinum pans support much higher temperatures, quartz pans are suitable for some open-pan liquid measurements, and stainless-steel pans are used for high-pressure work. The pan material must not react with the sample or cause unwanted effects within the measurement range.
-
Good sample preparation improves heat transfer and reproducibility. Powders should be distributed evenly and pressed gently into the pan. Solid blocks or pellets should have a flat surface to improve contact. Liquids should be placed in the center of the pan without contaminating the sealing area. Sheet materials should be cut to fit the pan, and multiple layers can be stacked when more mass is needed. Any material on the outside of the pan can contaminate the instrument and lead to measurement errors.
-
A larger sample produces a stronger signal and improves sensitivity, but it can broaden thermal events, shift some reaction temperatures higher, and reduce resolution between closely spaced events. A smaller sample improves resolution by narrowing transition intervals and helping separate overlapping peaks, but the signal may be weaker. The best practice is to use the smallest sample amount that still provides enough sensitivity for the measurement.
-
A faster heating rate increases peak sensitivity and can shorten measurement time, but it also shifts onset, peak, and offset temperatures higher. A slower heating rate improves resolution and helps separate overlapping thermal events. Atmosphere determines whether the sample reacts with its surroundings. In air, oxidation or combustion may occur; in nitrogen, the sample can be studied under inert conditions to observe behavior such as melting or decomposition without oxidation.
-
DSC can be applied to materials science, chemistry, pharmaceuticals, food science, and battery materials. It can track phase transitions in caffeine, monitor glass transition and crystallization changes in amorphous pharmaceutical compounds, identify melting and recrystallization behavior in polymorphic drugs, and characterize polymer layers in lithium-ion battery separators. These examples show its usefulness for studying stability, processing effects, safety behavior, and material performance.
Contact Us
Whether you're interested in getting a quote, want a demo, need technical support, or simply have a question, we're here to help.