UV Irradiation Unit

This attachment is suitable for the analysis of UV curing reactions such as in UV cured resins. The curing reaction during UV irradiation can be measured by DSC. It is equipped with two separate light guides allowing the sample and reference to be irradiated with UV light simultaneously while obtaining a stable baseline.

The UV light source unit can be selected from two types: LED lamp and mercury-xenon lamp, depending on the sample to be measured.

Specifications

UV light source unit

LED lamp

Mercury-xenon lamp

Measurement temperature range

Room temperature to 150℃

Irradiation wavelength range

365 nm, 405 nm *1

365 nm, 300 to 450 nm *2

Irradiation range

50 to 2000 mW/cm2 *3

2 to 100 mW/cm2

Power requirements

1Ø, AC100-240 V, 50/60 Hz, 0.6 A

1Ø, AC100-240 V, 50/60 Hz, 2.8 A

*1 Always replace the LED unit when changing the irradiation wavelength.

*2 Always remove the UV transmitting filter when changing the irradiation wavelength.

*3 Always install a condenser lens (accessory) when setting the irradiation wavelength to 400 mW/cm2 or more.

*4 Although a cooling unit can be connected, it does not support cooling measurements below room temperature.

✓ Cannot be used together with the sample observation unit.

✓ Cannot be used together with the gas selector.

Photo of UV Irradiation Unit

Contact Us

Whether you're interested in getting a quote, want a demo, need technical support, or simply have a question, we're here to help.