Titanium on Steel

    Application Note EDXRF1834

    EDXRF1834 banner

    Scope

    The measurement of titanium (Ti) conversion coating on galvanized steel is demonstrated.

    Background

    Aluminum and steel are often coated with a protective conversion coating, also called passivate or passivation coating, to prevent oxidation and corrosion of the base metal. Conversion coatings include chromium (Cr), titanium (Ti), vanadium (V), manganese (Mn), nickel (Ni), phosphorus (P), or zirconium (Zr). A phosphate coating may also be applied to minimize wear on cutting tools and stamping machines.

    Aluminum is often coated for use in aircraft parts, aluminum window frames, and other similar industries where the aluminum is exposed to weathering. Steel for
    the automotive industry is typically first galvanized with a zinc coating before the conversion coating is applied. Protected steel is also used for outdoor sheds and other similar uses where steel is exposed to weathering. Conversion coating also helps in the retention of paint for the final finished product.

    NEX QC+ high-resolution EDXRF analyzer

    Model: NEX QC

    Calibration

    An empirical calibration was built using a set of standards assayed by careful weigh-strip-weigh. 

    Element: Ti
    Units: mg/ft²
    Sample I.D. Standard value Calculated value
    STD 1 0.83 0.835
    STD 2 1.90 1.905
    STD 3 3.50 3.643
    STD 4 4.64 4.472
    STD 5 7.29 7.254
    STD 6 10.82 10.883
    STD 7 17.43 17.418

    EDXRF1834 Correlation plot Ti on galv steelCorrelation plot Ti on galvanized steel

    Recovery and repeatability

    To demonstrate repeatability (precision), the low and high calibration standards were chosen. Each sample was measured in a static position for ten repeat analyses with typical results shown below.

    Element: Ti
    Units: mg/ft²
    Sample I.D. Standard value Average value Std. dev % Relative
    STD 1 0.83 0.845 0.024 2.9
    STD 7 17.43 17.440 0.069 0.4

    Conclusion

    The performance shown here demonstrates NEX QC+ provides excellent sensitivity and performance for the measurement of titanium conversion coatings on galvanized steel.

    Contact Us

    Whether you're interested in getting a quote, want a demo, need technical support, or simply have a question, we're here to help.