Analysis of Lubricating Oils per ASTM D7751
Scope
Measurement of additive elements in lubricating oils by ASTM D7751 is demonstrated.
Background
Controlling additive elements when blending new, fresh lubricating oils is vital to ensure correct lubrication properties and minimizing production waste and product rejection. EDXRF is a fast and simple analytical technique for use by non-technical operators, yet powerful enough for expert use in the research & development of new lube oil formulations. To meet the industry needs in production and research Rigaku NEX CG offers compliance to D7751 Standard Test Method for Additive Elements in Lubricating Oils by EDXRF Analysis.
ASTM D7751-16
Table 1: Elements and range of applicability
Element | PLOQ in mass% | Max concentration in mass% |
Magnesium | 0.018 | 0.10 |
Phosphorous | 0.024 | 0.125 |
Sulfur | 0.008 | 1.94 |
Chlorine | 0.0007 | 0.05 |
Calcium | 0.002 | 0.44 |
Zinc | 0.040 | 0.143 |
Molybdenum | 0.004 | 0.047 |
NEX CG complies with ASTM D7751 when calibrated as per the method and good lab practices are followed.
Model: NEX CG
Calibration
Empirical calibrations were built using a suite of 23 commercially available calibration standards. The suite of calibration standards must be representative of the lube oil formulation to be analyzed. Elements in the lube oil should vary evenly over each concentration range of interest, and the elements in the oil should vary independently of each other. Empirical alpha corrections are then employed to automatically compensate for variations in X-ray absorption and enhancement effects within the sample due to the independent variations in element concentration. A summary of the empirical calibrations is shown here.
Element | Concentration range mass% |
Mg | 0.025 – 0.400 |
P | 0.010 – 0.350 |
S | 0.050 – 1.500 |
Cl | 0.0025 – 0.401 |
Ca | 0.025 – 1.000 |
Zn | 0.010 – 0.250 |
Mo | 0.0005 – 0.050 |
Precision
Two representative samples from the calibration suite were chosen to demonstrate typical instrument repeatability (precision). Ten repeat analyses of the sample were performed with the sample in static position.
Sample: ASI 5 Units: mass% |
||||
Element | Standard value | Average value | Std dev | % Relative |
Mg | 0.3000 | 0.2925 | 0.0052 | 1.7% |
P | 0.2000 | 0.2003 | 0.0007 | 0.4% |
S | 0.9500 | 0.9474 | 0.0075 | 0.8% |
Cl | 0.3754 | 0.3730 | 0.0075 | 2.0% |
Ca | 0.5500 | 0.5520 | 0.0017 | 0.3% |
Zn | 0.1600 | 0.1599 | 0.0007 | 0.4% |
Mo | 0.0140 | 0.0132 | 0.0007 | 5.0% |
Sample: ASI 16 Units: % |
||||
Element | Standard value | Average value | Std dev | % Relative |
Mg | 0.0250 | 0.0231 | 0.0039 | 16% |
P | 0.702 | 0.0701 | 0.0004 | 0.6% |
S | 0.5003 | 0.4932 | 0.0023 | 0.5% |
Cl | 0.0025 | 0.0020 | <0.0001 | <4% |
Ca | 0.6500 | 0.6545 | 0.0023 | 0.4% |
Zn | 0.2301 | 0.2326 | 0.0011 | 0.5% |
Conclusion
The Rigaku NEX CG combines indirect excitation with secondary targets, polarization targets and a high performance SDD detector to yield the optimum performance for elemental analysis of lube oils. Rigaku RPF-SQX fundamental parameters can also be used in association with matching libraries to encompass families of lubricants in a single method and for use with R&D testing new formulations making NEX CG an excellent tool for process and quality control in the manufacturing of lube oil formulations and blends as well as R&D and other uses.
Contact Us
Whether you're interested in getting a quote, want a demo, need technical support, or simply have a question, we're here to help.